R&S FSL  OUTPut Subsystem  
1300.2519.12 6.169  E-11 
OUTPut Subsystem 
The OUTPut subsystem controls the output features of the instrument.  
Commands of the OUTPut Subsystem 
– OUTPut<1|2>[:STATe] (models with tracking generator) 
– OUTPut<1|2>:IF[:SOURce] 
– OUTPut:UPORt[:VALue] 
– OUTPut:UPORt:STATe 
OUTPut<1|2>[:STATe] (models with tracking generator) 
 
This command switches the tracking generator on or off. 
Note:  With the tracking generator switched on, the maximum stop frequency is limited to 3 
GHz. This upper limit is automatically modified by the set frequency offset of the 
generator.  
If measurements in compliance with specs are to be performed with the tracking 
generator, the start frequency has to be 
3 x resolution bandwidth.  
If a sweep time shorter than recommended in the data sheet is selected, the sweep 
time indicator SWT on the screen is marked with a red asterisk and the message 
UNCAL is also displayed. 
With the tracking generator switched on, the FFT filters 
([SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]:TYPE) are not available. 
To switch off the tracking generator without loosing the the corresponding hardware 
settings and the normalization see the 
SOURce<1|2>:POWer[:LEVel][:IMMediate][:AMPLitude] command. 
Parameter 
ON | OFF 
Example 
OUTP ON 
Switches on the tracking generator. 
Characteristics 
*RST value: – 
SCPI: conform 
Mode 
A