OUTPut Subsystem R&S FSL
1300.2519.12 6.170 E-11
O
UTPut<1|2>:IF[:SOURce]
This command switches the source of the IF output between the demodulated signal (2 possible
values) and the IF signal.
Note: The AF output available at the frontpanel can only be used if the IF output source is set
to video.
Parameter
IF intermediate frequency output
VIDeo video output, 200 mV
HVIDeo video output, 1V
Example
OUTP IF VID
Selects the video signal for the IF output connector.
Characteristics
*RST value: IF
SCPI: conform
Mode
A
OUTPut:UPORt[:VALue]
This command sets the control lines of the user ports.
The user port is written to with the given binary pattern. If the user port is programmed to INPut
instead of OUTPut, the output value is temporarily stored.
Parameter
#B00000000 to #B11111111
Example
OUTP:UPOR #B10100101
Characteristics
*RST value: –
SCPI: device–specific
Mode
all