EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Status:operation Register; Status:questionable Register

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSL Status Reporting System
1300.2519.12 5.19 E-11
STATus:OPERation Register
In the CONDition part, this register contains information on which actions the instrument is being
executing or, in the EVENt part, information on which actions the instrument has executed since the last
reading. It can be read using the commands STATus:OPERation:CONDition? or STATus
:OPERation[:EVENt]?.
Table 5–5 Meaning of the bits in the STATus:OPERation register
Bit No.
Meaning
0
CALibrating
This bit is set as long as the instrument is performing a calibration.
1 to 7
Not used
8
HardCOPy in progress
This bit is set while the instrument is printing a hardcopy.
9 to 14 Not used
15 This bit is always 0
STATus:QUEStionable Register
This register contains information about indefinite states which may occur if the unit is operated without
meeting the specifications. It can be read using the commands STATus:QUEStionable:
CONDition? and STATus:QUEStionable[:EVENt]?.
Table 5–6 Meaning of bits in STATus:QUEStionable register
Bit No. Meaning
0 to 1
These bits are not used
2
BATTERY LOW
If the instrument is running without any external power supply and the charging level of the
internal battery is approximately lower than 5% this bit is set to indcate that the system will
be shut down automatically in approx. 5 minutes.
3
POWer
This bit is set if a questionable power occurs (refer also to section
"STATus:QUEStionable:POWer Register")
4
TEMPerature
This bit is set if a questionable temperature occurs.
5
FREQuency
The bit is set if a frequency is questionable (refer to section
"STATus:QUEStionable:FREQuency Register")
6 to 7
Not used

Table of Contents