CALibration Subsystem (Power Meter, K9) R&S FSL
1300.2519.12 6.380 E-11
CALibration Subsystem (Power Meter, K9)
The CALibration subsystem determines the data for system error correction in the instrument.
The following subsystem is included:
• "CALibration:PMETer Subsystem (Power Meter, K9)" on page 6.380
CALibration:PMETer Subsystem (Power Meter, K9)
The CALibration:PMETer Subsystem determines the error correction data for measurements in the
power meter mode.
Commands of the CALibration:PMETer Subsystem
– CALibration:PMETer:ZERO:AUTO
CALibration:PMETer:ZERO:AUTO
This command starts zeroing of the power sensor.
Parameter
ONCE
Example
:CAL:PMET:ZERO:AUTO ONCE;*WAI
Starts zeroing and delays the execution of further commands until zeroing is concluded.
Characteristics
*RST value: –
SCPI: device–specific
Mode
PSM