R&S FSL TRIGger Subsystem
1300.2519.12 6.277 E-11
TRIGger Subsystem
The TRIGger subsystem is used to synchronize instrument actions with events. It is thus possible to control
and synchronize the start of a sweep.
Commands of the TRIGger Subsystem
TRIGger<1|2>[:SEQuence]:HOLDoff[:TIME]
TRIGger<1|2>[:SEQuence]:IFPower:HOLDoff
TRIGger<1|2>[:SEQuence]:IFPower:HYSTeresis
TRIGger<1|2>[:SEQuence]:LEVel:IFPower
TRIGger<1|2>[:SEQuence]:LEVel:VIDeo
TRIGger<1|2>[:SEQuence]:SLOPe
TRIGger<1|2>[:SEQuence]:SOURce
TRIGger<1|2>[:SEQuence]:TIME:RINTerval
TRIGger<1|2>[:SEQuence]:VIDeo:CONTinuous (option TV Trigger, B6)
TRIGger<1|2>[:SEQuence]:VIDeo:FIELd:SELect (option TV Trigger, B6)
TRIGger<1|2>[:SEQuence]:VIDeo:FORMat:LPFRame (option TV Trigger, B6)
TRIGger<1|2>[:SEQuence]:VIDeo:LINE:NUMBer (option TV Trigger, B6)
TRIGger<1|2>[:SEQuence]:VIDeo:SSIGnal:POLarity (option TV Trigger, B6)
TRIGger<1|2>[:SEQuence]:HOLDoff[:TIME]
This command defines the length of the trigger delay.
A negative delay time (pretrigger) can be set in zero span only.
Parameter
–100 s to +100 s
for TV trigger (option B6, available from firmware version 1.10): –50 Us to +50 U s
Example
TRIG:HOLD 500us
Characteristics
*RST value: 0 s
SCPI: conform
Mode
all