TRIGger Subsystem R&S FSL
1300.2519.12 6.280 E-11
T
RIGger<1|2>[:SEQuence]:SLOPe
This command selects the slope of the trigger signal. The selected trigger slope applies to all
trigger signal sources except the TV trigger (option B6, available from firmware version 1.10).
Parameter
POSitive | NEGative
Example
TRIG:SLOP NEG
Characteristics
*RST value: POSitive
SCPI: conform
Mode
all
TRIGger<1|2>[:SEQuence]:SOURce
This command selects the trigger source for the start of a sweep.
Parameter
IMMediate (Free Run) | EXTern | IFPower | VIDeo | TIME (available from firmware version 1.60)
TV (option TV Trigger, B6, available from firmware version 1.10)
For details on trigger modes refer to chapter "Instrument Functions", section "Trigger mode
overview".
Example
TRIG:SOUR EXT
Selects the external trigger input as source of the trigger signal
Characteristics
*RST value: IMMediate
SCPI: conform
Mode
A, ADEMOD, CATV, CDMA, EVDO, SPECM, WCDMA
TRIGger<1|2>[:SEQuence]:TIME:RINTerval
This command sets the time intervall for the time trigger source.
The numeric suffixes <1|2> are irrelevant for this command.
Parameter
0.1 to 5000 s
Example
TRIG:SOUR TIME
Selects the time trigger input for triggering.
TRIG:TIME:RINT 50
The sweep starts after 50 s.