Noise Figure Measurements Option (K30)  R&S FSL  
1300.2519.12 2.104  E-11 
Fig. 2-75:  Calibration and measurement on DUTs with a high gain 
Frequency–Converting Measurements 
The frequency–converting measurement is used for DUTs with an output frequency that differs from the 
input frequency, e.g. mixers and frequency converters. The frequency–converting measurement allows 
many variations, which differ from each other in two criteria: 
 Fixed LO Measurements 
 Image–Frequency Rejection (SSB, DSB) 
Fixed LO Measurements 
In the Frequency Settings dialog box, select one of the following settings for the Mode parameter: 
 fixed LO, IF=RF+LO, for up–converting devices 
 fixed LO, IF=abs(RF–LO), for down converters or image measurements 
 
Image–Frequency Rejection (SSB, DSB) 
Frequency–converting DUTs often do not only convert the desired input frequency but also the image 
frequency. A broadband noise source offers noise to the DUT not only at the input frequency but also at 
the image frequency. If the noise power at the IF gate is measured, the origin of the noise can no longer 
be determined. It may have been converted both from the input and from the image frequency range. 
Test setup 
 Set the following parameters: 
 IF (intermediate frequency): 100 MHz 
 RF (input frequency): 400 MHz 
 LO (local oscillator frequency): 500 MHz 
 image (image frequency): 600 MHz