EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Configure Subsystem (CDMA, K82)

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands of the CDMA2000 BTS Analyzer Option (K82) R&S FSL
1300.2519.12 6.604 E-11
CONFigure Subsystem (CDMA, K82)
The CONFigure subsystem contains commands for configuring the measurements.
The following subsystem is included:
"CONFigure:CDPower Subsystem (CDMA, K82)" on page 6.604
CONFigure:CDPower Subsystem (CDMA, K82)
This subsystem contains the commands for measurement selection and configuration of the
CDMA2000 BTS Analyzer option.
Further setting commands for the spectrum emission mask measurement are in the
"CALCulate:LIMit:ESPectrum Subsystem (CDMA, K82)" on page 6.599.
Commands of the CONFigure:CDPower Subsystem
CONFigure:CDPower[:BTS]:CTABle[:STATe]
CONFigure:CDPower[:BTS]:CTABle:CATalog?
CONFigure:CDPower[:BTS]:CTABle:COMMent
CONFigure:CDPower[:BTS]:CTABle:COPY
CONFigure:CDPower[:BTS]:CTABle:DATA
CONFigure:CDPower[:BTS]:CTABle:DELete
CONFigure:CDPower[:BTS]:CTABle:NAME
CONFigure:CDPower[:BTS]:CTABle:SELect
CONFigure:CDPower[:BTS]:BCLass
CONFigure:CDPower[:BTS]:MCARier[:STATe]
CONFigure:CDPower[:BTS]:MEASurement
CONFigure:CDPower:CTABle:RESTore
CONFigure:CDPower[:BTS]:CTABle[:STATe]
This command activates or deactivates the ‘RECENT’ channel table. To select another channel
table, use the CONFigure:CDPower[:BTS]:CTABle:SELect command.
Parameter
ON | OFF
Example
INST:SEL BC2K
Activates the CDMA2000 BTS Analyzer option.
INIT:CONT OFF
Selects single sweep.
INIT;*WAI
Starts measurement with synchronization.
CONF:CDP:CTAB ON
Activates predefined channel table.
CONF:CDP:CTAB:SEL 'CTAB_1'
Selects channel table.

Table of Contents