R&S FSL CONFigure Subsystem (CDMA, K82)
1300.2519.12 6.609 E-11
C
ONFigure:CDPower[:BTS]:MCARier[:STATe]
This command activates or deactivates the settings that improve the processing of multi-carrier
signals.
Parameter
ON | OFF
Example
INST:SEL BC2K
Activates the CDMA2000 BTS Analyzer option.
INIT:CONT OFF
Selects single sweep.
CONF:CDP:MCAR ON
Activates the multi-carrier settings.
Characteristics
RST value: OFF
SCPI: device-specific
Mode
CDMA
CONFigure:CDPower[:BTS]:MEASurement
This command selects the measurements of the CDMA2000 BTS Analyzer option. The
predefined settings of the different measurements are described in chapter 4, section "Softkeys
of the measurement menu (CDMA2000 BTS Analyzer mode)".Parameter
ACLR Adjacent–Channel Power measurement with predefined settings
according to the CDMA2000 standard
CCDF measurement of the complementary cumulative distribution
function (signal statistics)
CDPower Code Domain Analyzer measurement.
ESPectrum check of signal power (Spectrum Emission Mask)
OBWidth measurement of the occupied bandwidth
POWer Signal Channel Power measurement with predefined settings
according to the CDMA2000 standard
Example
INST:SEL BC2K
Activates the CDMA2000 BTS Analyzer option.
INIT:CONT OFF
Selects single sweep.
CONF:CDP:MEAS POW
Selects Signal Channel Power measurement .
INIT;*WAI
Starts measurement with synchronization.

Table of Contents