CALCulate Subsystem (Power Meter, K9) R&S FSL
1300.2519.12 6.378 E-11
CALCulate Subsystem (Power Meter, K9)
The CALCulate subsystem contains commands for converting instrument data, transforming and
carrying out corrections. These functions are carried out subsequent to data acquisition, i.e. following
the SENSe subsystem.
The following subsystems are included:
"CALCulate:PMETer Subsystem" on page 6.378
CALCulate:PMETer Subsystem (Power Meter, K9)
This subsystem controls the instrument settings for power sensor measurements.
Commands of the CALCulate:PMETer Subsystem
CALCulate<1|2>:PMETer:RELative[:MAGNitude]
CALCulate<1|2>:PMETer:RELative[:MAGNitude]:AUTO
CALCulate<1|2>:PMETer:RELative:STATe
CALCulate<1|2>:PMETer:RELative[:MAGNitude]
This command sets the reference value for relative measurements.
The numeric suffixes <1|2> are irrelevant.
Parameter
–200 to 200
Example
CALC:PMET:REL –30
Sets the reference value for relative measurements to –30 dBm.
Characteristics
*RST value: 0
SCPI: device–specific
Mode
PSM
CALCulate<1|2>:PMETer:RELative[:MAGNitude]:AUTO
This command takes the current measurement value as reference value for relative
measurements.
The numeric suffixes <1|2> are irrelevant.
Parameter
ONCE
Example
CALC:PMET:REL:AUTO ONCE
Takes the current measurement value as reference value for relative measurements.