TRIGger Subsystem R&S FSL
1300.2519.12 6.278 E-11
T
RIGger<1|2>[:SEQuence]:IFPower:HOLDoff
ffThis command sets the holding time before the next IF power trigger event.
T
his command is available from firmware version 1.30.
Parameter
<numeric_value> in s: 150 ns to 1000 s
Example
TRIG:SOUR IFP
Sets the IF power trigger source.
TRIG:IFP:HOLD 200 ns
Sets the holding time to 200 ns.
Characteristics
*RST value: 150 ns
SCPI: device–specific
Mode
A–F
TRIGger<1|2>[:SEQuence]:IFPower:HYSTeresis
sThis command sets the limit that the hysteresis value has to fall below in order to trigger the
next measurement.
This command is available from firmware version 1.30.
Parameter
<numeric_value> in dB: 3 dB to 50 dB
Example
TRIG:SOUR IFP
Sets the IF power trigger source.
TRIG:IFP:HYST 10DB
Sets the hysteresis limit value.
Characteristics
*RST value: 3 dB
SCPI: device–specific
Mode
A–F