DISPlay Subsystem R&S FSL
1300.2519.12 6.130 E-11
D
ISPlay:PSAVe[:STATe]
This command switches on or off the power-save mode of the display. With the power-save
mode activated the display including backlight is completely switched off after the elapse of the
response time (see DISPlay:PSAVe:HOLDoff command).
Note: This mode is recommended for preserving the display especially if the instrument is
exclusively operated via remote control.
This command is available from firmware version 1.20.
Parameter
ON | OFF
Example
DISP:PSAVe ON
Switches on the power-save mode.
Characteristics
*RST value: OFF
SCPI: device-specific
Mode
all
DISPlay:PSAVe:HOLDoff
This command sets the holdoff time for the power-save mode of the display. The available value
range is 1 to 60 minutes, the resolution 1 minute. The entry is dimensionless.
This command is available from firmware version 1.20.
Parameter
1 to 60
Example
DISP:PSAV:HOLD 30
Characteristics
*RST value: 15
SCPI: device-specific
Mode
all
DISPlay[:WINDow<1|2>]:SIZE
This command configures the measurement display.
Channel and adjacent–channel power measurements: Only "1" is allowed as a numeric suffix.
Spectrum Emission Mask and Spurious Emissions measurements: The numeric suffix <1|2>
selects the item that is displayed in full size; 1 for measurement diagram, 2 for list
Parameter
LARGe Channel and adjacent–channel power measurements: diagram in full screen.
Spectrum Emission Mask and Spurious Emissions measurements: diagram or
list in full screen, depending on the value of the numeric suffix.
SMALl split screen (diagram and list)