DISPlay Subsystem R&S FSL
1300.2519.12 6.136 E-11
C
haracteristics
*RST value: 0dB
SCPI: conform
M
ode
A, CATV, CDMA, EVDO, PSM, WCDMA, WLAN
DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:RPOSition (models with tracking generator)
This command requires a tracking generator and active normalization in the Tracking
Generator mode. It defines the position of the reference value.
The numeric suffixes <1|2> and <1...6> are irrelevant.
Parameter
0 to 100PCT
Example
DISP:TRAC:Y:RPOS 50PCT
Characteristics
*RST value:
100 PCT Spectrum Analyzer mode
50 PCT Tracking Generator mode
SCPI: conform
Mode
A, ADEMOD, BT, CATV, CDMA, EVDO, WCDMA
DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:RVALue (models with tracking generator)
This command requires a tracking generator and active normalization in the Tracking
Generator mode. The command defines the power value assigned to the reference position.
The numeric suffixes <1|2> and <1...6> are irrelevant.
Parameter
<numeric_value>
Example
DISP:TRAC:Y:RVAL –20dBm
(Analyzer)
DISP:TRAC:Y:RVAL 0
Sets the power value assigned to the reference position to 0 dB (tracking generator)
Characteristics
*RST value: 0 dB, coupled to reference level
SCPI: device–specific