R&S FSL CALibration Subsystem
1300.2519.12 6.121 E-11
C
ALibration:ABORt
This command has no effect, but is implemented to provide compatibility with the FSP family.
The collection of self alignment data can only be aborted via manual operation. For details refer
to chapter 4 "Instrument Functions - Basic Settings", section "Instrument Setup and Interface
Configuration - SETUP Key" - "Alignment".
This command is an event and therefore has no *RST value and no query.
Example
CAL:ABOR
Characteristics
*RST value: -
SCPI: device-specific
Mode
all
CALibration:RESult?
This command outputs the results of the correction data acquisition. The lines of the result table
are output as string data separated by commas:
Total Calibration Status: PASSED,Date (dd/mm/yyyy): 12/07/2004,
Time: 16:24:54,Runtime: 00.06
Example
CAL:RES?
Characteristics
*RST value: -
SCPI: device-specific
Mode
all
CALibration:STATe
This command determines whether the current calibration data are taken into account by the
instrument (ON) or not (OFF).
Parameter
ON | OFF
Example
CAL:STAT OFF
Sets up the instrument to ignore the calibration data.
Characteristics
*RST value: -
SCPI: conform
Mode
all