R&S FSL Tracking Generator (Models 13, 16 and 28)
1300.2519.12 4.137 E-11
Softkeys of the tracking generator menu
The following table shows all softkeys available in the tracking generator menu. It is possible that your
instrument configuration does not provide all softkeys. If a softkey is only available with a special option,
model or (measurement) mode, this information is delivered in the corresponding softkey description.
For the description of the other main softkeys refer to "Optional softkeys of the menu menu" on page
4
.131.
Menu / Command Command
Source On/Off
Source Power
Power Offset
Source Cal Cal Trans
Cal Refl Short
Cal Refl Open
Normalize
Ref Value Position
Ref Value
Recall
Source On/Off
Switches the tracking generator on or off. Default setting is off.
If the tracking generator is switched off, the corresponding hardware settings and the
normalization are discarded. To switch off the tracking generator but keep the hardware settings
and the normalization, enter –400 dBm into the edit dialog box displayed by pressing the
Source Power softkey.
Remote: OUTP:STAT ON
Source Power
Opens an edit dialog box to enter a tracking generator output power. The default output power is
–20 dBm. The range is specified in the data sheet. Additionally, the value –400 dBm is available
to switch off the tracking generator but keep the hardware settings.
If the tracking generator is off, it is automatically switched on if an output power value is entered.
For details on switching on or off refer to the Source On/Off softkey.
Remote: SOUR:POW –20dBm