Tracking Generator (Models 13, 16 and 28) R&S FSL
1300.2519.12 4.138 E-11
Power Offset
Opens an edit dialog box to enter a constant level offset for the tracking generator. Values from
–200 dB to +200 dB in 1 dB steps are allowed. The default setting is 0 dB. Offsets are indicated
by the enhancement label LVL.
With this offset for example attenuators or amplifiers at the output connector of the tracking
generator can be taken into account for the displayed output power values on screen or during
data entry. Positive offsets apply to an amplifier and negative offsets to an attenuator
subsequent to the tracking generator.
Remote: SOUR:POW:OFFS –10dB
Source Cal
Opens a submenu to configure calibration for transmission and reflection measurement. For
details on the test setups see "Transmission measurement" on page 4.135 and "Reflection
measurement" on page 4.135.
Cal Trans
Starts a sweep that records a reference trace. This trace is used to calculate the difference for
the normalized values.
Remote: CORR:METH TRAN
Cal Refl Short
Starts a sweep as reference trace for short–circuit calibration.
If both calibrations (open circuit, short circuit) are carried out, the calibration curve is calculated
by averaging the two measurements and stored in the memory. The order of the two calibration
measurements is free.
Remote: CORR:METH REFL
Cal Refl Open
Starts a sweep as reference trace the open–circuit calibration.
If both calibrations (open circuit, short circuit) are carried out, the calibration curve is calculated
by averaging the two measurements and stored in the memory. The order of the two calibration
measurements is free.
Remote: CORR:COLL OPEN
Normalize
Switches the normalization on or off. The softkey is only available if the memory contains a
reference trace. For details on normalization see "Calibration mechanism" on page 4.136.
Remote: CORR ON