EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 1420

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSL SENSe Subsystem (EVDO, K84)
1300.2519.12 6.661 E-11
[
SENSe<1|2>:]CDPower:OVERview
This command can be enabled can be precisely enabled by means of ON when either the Code
Domain Power or the Code Domain Error Power analysis is active (refer to
CALCulate<1|2>:FEED). In Overview mode, the I branch of the signal is normally displayed on
screen A and the Q branch of the signal on screen B with the CDP/CDEP. The branches can be
read out separately by means of TRAC:DATA? TRACE1 and TRAC:DATA? TRACE2. The
previous analyses become active again when you exit Overview mode.
If an analysis other than Code Domain Power or Code Domain Error Power is selected when
Overview mode is active, you exit Overview mode and the previous analysis is reset on the
other screen.
Parameter
ON | OFF
Example
CDP:OVER ON
Activate Overview mode.
CDP Relative on screen A I branch.CDP Relative on screen B Q branch
Characteristics
RST value: OFF
SCPI: device-specific
Mode
EVDO
[SENSe<1|2>:]CDPower:PNOFfset
This command sets the PN offset of the base station in multiples of 64 chips.
The numeric suffix at SENSe<1|2> is irrelevant for this command.
Parameter
0 to 511
Example
INST:SEL BC2K
Activates the 1XEV-DO BTS Analyzer option.
INIT:CONT OFF
Selects single sweep.
CDP:PNOF 45
Sets PN offset.
INIT;*WAI
Starts measurement with synchronization.
Characteristics
RST value: 0
SCPI: device-specific
Mode
EVDO

Table of Contents