R&S FSL SENSe Subsystem (EVDO, K84)
1300.2519.12 6.657 E-11
Characteristics
RST value: 0
SCPI: device-specific
Mode
EVDO
[SENSe<1|2>:]CDPower:CODE
This command selects the code number. The maximum number depends on the channel type
The numeric suffix at SENSe<1|2> is irrelevant for this command.
Parameter
Channel Type Spreading factor Code number
PILOT 32 031
MAC REV 0 64
REV A 128
0...63
0...127
PREAMBLE REV 0 32
REV A 64
0...31
0...63
DATA 64 0...31
Example
CDP:CODE 11
Selects code number 11.
Characteristics
RST value: 0
SCPI: device-specific
Mode
EVDO
[SENSe<1|2>:]CDPower:ICTReshold
This command defines the minimum power which a single channel must have compared to the
total signal in order to be regarded as an active channel. Channels below the specified
threshold are regarded as "inactive".
The numeric suffix at SENSe<1|2> is irrelevant for this command.
Parameter
–100 dB to 0 dB
Example
CDP:ICTR –10DB
Sets threshold value to –10 dB.
Characteristics
RST value: –40 dB
SCPI: device-specific
Mode
EVDO