SENSe Subsystem (Power Meter, K9) R&S FSL
1300.2519.12 6.386 E-11
[SENSe<1|2>:]PMETer:ROFFset[:STATe]
This command defines whether the reference level offset set for the analyzer is taken into
account for the measured power or not.
The numeric suffixes <1|2> are irrelevant.
This command is available from firmware version 1.50.
Parameter
ON Adds the offset defined by the
DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:RLEVel:OFFSet
command to the measured level.
OFF Takes no offset into account.
Example
PMET:ROFF OFF
Takes no offset into account for the measured power.
Characteristics
*RST value: ON
SCPI: device–specific
Mode
PSM
[SENSe<1|2>:]PMETer:TRIGger[:STATe]
This command switches the external power trigger on or off.
The numeric suffixes <1|2> are irrelevant.
This command is only available in conjunction with a NRP-Z81 power sensor.
This command is available from firmware version 1.90.
Parameter
ON | OFF
Example
PMET:TRIG ON
Switches the external power trigger on.
Characteristics
*RST value: OFF
SCPI: device–specific
Mode
PSM