SENSe Subsystem (Analog Demodulation, K7) R&S FSL
1300.2519.12 6.308 E-11
INIT;*WAI
Starts the measurement and waits for the termination
FORM ASC
Selects the output format
ADEM:FM:RES? AVER
Reads the FM average result data
ADEM:FM:RES? MAXH
Reads the FM Maxhold result data
ADEM:FM:RES? MINH
Reads the FM Minhold result data
ADEM:AM:REL:RES? WRIT
Reads the current AM result data
ADEM:AM:REL:AFSP:RES? WRIT
Reads the current AF spectrum result data of the demodulated AM signal
Characteristics
*RST value: –
SCPI: device–specific
Mode
ADEMOD
[SENSe:]ADEMod:BANDwidth|BWIDth:DEModulation
This command defines the demodulation bandwidth used for analog demodulation. The required
sampling rate is automatically set depending on the selected demodulation bandwidth. The
available demodulation bandwidths are determined by the existing sampling rates (see table
below).
Parameter
<numeric_value>
For details on the correlation of bandwidth and sample rate refer to chapter "Instrument
Functions", section "Analog Demodulation (Option K7)" – "Sample rate, measurement time and
trigger offset".
Example
ADEM:BAND:DEM 1MHz
Sets the demodulation bandwidth to 1 MHz.
Characteristics
*RST value: 5 MHz
SCPI: device–specific
Mode
ADEMOD