SENSe Subsystem (CATV, K20) R&S FSL
1300.2519.12 6.488 E-11
[SENSe<1|2>:]ATV:CN:POWer:REFerence:MODE
This command defines the reference power method.
The numeric suffixes <1|2> are irrelevant.
Parameter
RCHannel reference channel
([SENSe<1|2>:]ATV:CN:POWer:REFerence:CHANnel:MANual
command)
MCHannel same as measurement channel
MANual manual reference power
([SENSe<1|2>:]ATV:CN:POWer:REFerence:MANual command)
Example
ATV:CN:POW:REF:MODE RCH
Sets the reference channel method.
Characteristics
RST value: MCH
SCPI: device–specific
Mode
CATV
[SENSe<1|2>:]ATV:CN:TABle<1...10>:MFRequencies
This command defines the measurement frequencies (center frequencies and span values), and
activates or deactivates them for the C/N measurement.
The numeric suffix <1...10> defines the table row. The numeric suffixes <1|2> are irrelevant.
Parameter
0 | 1,<numeric_value>,<numeric_value>
0 deactivated
1 activated
first numeric value center frequency in Hz
second numeric value span in Hz
Example
ATV:CN:TABL1:MFR 1,1000,1000
Activates the entries of the first table row for the measurement and sets the center frequency to
1 kHz and the span to 1 kHz.
ATV:CN:TABL2:MFR 1,2000,2000
Activates the second row entries for the measurement and sets the center frequency to 2 kHz
and the span to 2 kHz.
Characteristics
RST value: 1,1.25 KHz,1 MHz (row 1); 0,1.25 kHz,1 MHz (row 2–10)
SCPI: device–specific
Mode