R&S FSL Measuring the Channel and Adjacent Channel Power
1300.2519.12 7.11 E-11
'––––––––– Read out result –––––––––––––––––––––––––––––––––––––––––––––––––
IF (status% = 1) THEN
CALL InstrWrite(analyzer,"STAT:QUES:LIM1:COND?")
'Read out STAT:QUES:LIMit
'
register
CALL InstrRead(analyzer, result$)
IF ((Val(result$) And 16) <> 0) THEN
Print "Limit5 failed"
ELSE
Print "Limit5 passed"
END IF
END IF
END SUB
REM
************************************************************************
Measuring the Channel and Adjacent Channel Power
In the following example, the channel and adjacent channel power is first measured on a signal with a
level of 0 dBm at 800 MHz to IS95. Then the channel and adjacent channel power is measured on a
GSM signal at 935.2 MHz with fast ACP measurement (FAST ACP).
In addition, the limit test is activated.
REM ************************************************************************
Public Sub ACP()
result$ = Space$(100)
'––––––––– R&S FSL default setting ––––––––––––––––––––––––––––––––––––––––
CALL SetupStatusReg 'Set status register
CALL InstrWrite(analyzer,"*RST") 'Reset instrument
CALL InstrWrite(analyzer,"INIT:CONT OFF") 'Single sweep
CALL InstrWrite(analyzer,"SYST:DISP:UPD ON")
'ON: display on
'OFF: off
'––––––––– Set frequency –––––––––––––––––––––––––––––––––––––––––––––––––––
CALL InstrWrite(analyzer,"FREQ:CENT 800MHz")
'Set frequency
'––––––––– Set level –––––––––––––––––––––––––––––––––––––––––––––––––––––––
CALL InstrWrite(analyzer,"DISP:WIND:TRAC:Y:RLEV 10dBm")
'Reference level
'––––––––– Example 1: Configure CP/ACP for CDMA––––––––––––––––––––––––––––
CALL InstrWrite(analyzer,"CALC:MARK:FUNC:POW:SEL ACP")
'ACP measurement on
CALL InstrWrite(analyzer,"CALC:MARK:FUNC:POW:PRES F8CDMA")
'Select CDMA800 FWD
CALL InstrWrite(analyzer,"SENS:POW:ACH:ACP 2")
'Select 2 adjacent channels