Spectrum Emission Mask Measurement R&S FSL
1300.2519.12 7.28 E-11
'––––––––– Defining the reference range settings –––––––––––––––––––––––––––
Rem The bandwith of the reference range limits the minimum span
Rem of the reference range definition later in the script.
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RTYPe CPOWer")
'
Set power reference type
C
ALL InstrWrite(analyzer,":SENSe1:ESPectrum:BWID 2 MHZ")
'Set bandwidth
'––––––––– Defining the settings of range 1 ––––––––––––––––––––––––––––––––
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe1:FREQuency:STARt –4MHz")
'Set the start frequency
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe1:FREQuency:STOP –2MHz")
'Set the stop frequency
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe1:BANDwidth:RESolution
1 MHZ")
'Set the resolution bandwidth
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe1:FILTer:TYPE CFILter")
'Set the channel filters
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe1:BANDwidth:VIDeo 3 MHZ")
'Set the video bandwidth to 3 MHz
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe1:SWEep:TIME 20 ms")
'Set the sweep time to 20 ms
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe1:RLEVel 5 DBM")
'Set the reference level to 5 dBm
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe1:INPut:ATTenuation 20 DB")
'Set the attenuation to 20 dB
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe1:LIMit:ABSolute:STARt
–50")
'Set an absolute limit of –50 dBm
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe1:LIMit:ABSolute:STOP –50")
'Set an absolute limit of –50 dBm
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe1:LIMit:RELative:STARt
–70")
'Set a relative limit of –70 dBc
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe1:LIMit:RELative:STOP –60")
'Set a relative limit of –60 dBc
'––––––––– Defining the settings of range 2 ––––––––––––––––––––––––––––––––
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe2:FREQuency:STARt –2MHz")
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe2:FREQuency:STOP –1MHz")
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe2:BANDwidth:RESolution
100 kHz")
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe2:FILTer:TYPE NORM")
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe2:BANDwidth:VIDeo 300 kHZ")
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe2:SWEep:TIME 50 ms")
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe2:RLEVel 10 DBM")
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe2:INPut:ATTenuation 30 DB")
CALL InstrWrite(analyzer,":SENSe1:ESPectrum:RANGe2:LIMit:ABSolute:STARt
–40")