R&S FSL Device–Specific Messages
1300.2519.12 9.7 E-11
Device–Specific Messages
Error code Error text in the case of queue poll
Error explanation
1052
Frontend LO is Unlocked
This message is displayed when the phase regulation of the local oscillator fails in the RF front–end.
1060
Trigger–Block Gate Delay Error– gate length < Gate Delay
This message is displayed when the gate signal length is not sufficient for the pull–in delay with a
predefined gate delay.
1064
Tracking LO is Unlocked
This message is displayed when the phase regulation of the local oscillator fails on the tracking generator
module.
2028
Hardcopy not possible during measurement sequence
This message is displayed when a printout is started during scan sequences that cannot be interrupted.
Such sequences are for example:
Recording the system error correction data (alignment)
Instrument self test
In such cases synchronization to the end of the scan sequence should be performed prior to starting the
printout.
2033
Printer Not Available
This message is displayed when the selected printer is not included in the list of available output devices.
A possible cause is that the required printer driver is missing or incorrectly installed.
2034
CPU Temperature is too high
This message is displayed when the temperature of the processor exceeds 70 °C.