EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 220

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSL Triggering the Sweep – TRIG Key
1300.2519.12 4.35 E-11
Repetition Intervall (Time Trigger mode)
Opens an edit dialog box to enter the time intervall after which the sweep is started. The
possible values range from 100 ms to 5000 s.
This softkey is available from firmware version 1.60.
Remote: TRIG:TIME:RINT 50
Gated Trigger (option Gated Sweep, B8)
Switches the sweep mode with gate on or off .
This softkey requires the following trigger mode:
span > 0
External or IF Power
span = 0
External or IF Power or Video
If a different mode is active, the IF Power trigger mode is automatically selected.
If the gate is switched on, a gate signal applied to the rear panel connector EXT
TRIGGER/GATE or the internal IF power detector controls the sweep of the analyzer.
In the Time Trigger mode, this softkey is not available.
For details also see "To use gated sweep operation (option Gated Sweep, B8)" on page 4.29.
Remote: SWE:EGAT ON
Remote: SWE:EGAT:SOUR IFP | EXT
Gate Settings (option Gated Sweep, B8)
Opens a submenu to make all the settings required for gated sweep operation.
In the Time Trigger mode, this softkey is not available.
For details also see "To use gated sweep operation (option Gated Sweep, B8)" on page 4.29.
Gate Mode Lvl/Edge (option Gated Sweep, B8)
Sets the gate mode. As settings level–triggered or edge–triggered gate mode can be selected.
For details also see "To use gated sweep operation (option Gated Sweep, B8)" on page 4.29.
Remote: SWE:EGAT:TYPE EDGE