R&S FSL Measurement of Harmonics
1300.2519.12 2.5 E-11
The mode changes from phase noise measurement to reference fixed, the marker readout
c
hanges from dB/Hz to dB.
Fig. 2-2 Fundamental wave and the frequency and level reference point
5. Make the step size for the center frequency equal to the signal frequency
Press the FREQ key.
The frequency menu is displayed.
Press the CF–Stepsize softkey and press the = Marker softkey in the submenu.
The step size for the center frequency is now equal to the marker frequency.
6. Set the center frequency to the 2
nd
harmonic of the signal
Press the FREQ key.
The frequency menu is displayed.
Press the UPARROW key once.
The center frequency is set to the 2
nd
harmonic.
7. Place the delta marker on the 2
nd
harmonic.
Press the MKR–>
key.
Press the Peak softkey.