R&S FSL Power Measurements – MEAS Key
1300.2519.12 4.117 E-11
Harmonic Sweep Time
For details refer to the Sweeptime Manual softkey in the bandwidth menu.
This softkey is available from firmware version 1.10.
Harmonic RBW Auto
Enables/disables the automatic adjustment of the resolution bandwidth. The automatic
adjustment is carried out according to:
RBWn = RBW
1
* n
If RBW
n
is not available, the next higher value is used.
This softkey is available from firmware version 1.10.
Remote: CALC:MARK:FUNC:HARM:BAND:AUTO OFF
Adjust Settings
Activates the frequency search in the frequency range that was set before starting the harmonic
measurement (if harmonic measurement was with span > 0) and adjusts the level.
This softkey is available from firmware version 1.10.
Remote: CALC:MARK:FUNC:HARM:PRES
Full Size Diagram (span > 0)
Displays the diagram in full screen size.
Remote: DISP:SIZE LARG|SMAL