1xEV-DO BTS Analyzer (Option K84) R&S FSL
1300.2519.12 4.398 E-11
Marker 1/Marker 2/Marker 3/Marker 4/Marker Norm/Delta
The Marker <no> softkey activates the corresponding marker and opens an edit dialog box to
enter a value for the marker to be set to. Pressing the softkey again deactivates the selected
marker. If two screens are active in the Code Domain Analyzer, each of the screens has its own
set of markers.
Marker 1 is always the reference marker for relative measurements. If activated, markers 2 to 4
are delta markers that refer to marker 1. These markers can be converted into markers with
absolute value display by means of the Marker Norm/Delta softkey. If marker 1 is the active
marker, pressing the Marker Norm/Delta softkey switches on an additional delta marker.
For the Channel Bitstream measurement only one marker (Marker 1) is available. It can be used
for scrolling and to display the number and value of a bit.
Remote: CALC:MARK ON
Remote: CALC:MARK:X <value>
Remote: CALC:MARK:Y?
Remote: CALC:DELT ON
Remote: CALC:DELT:X <value>
Remote: CALC:DELT:X:REL?
Remote: CALC:DELT:Y?
All Marker Off
For details refer to the All Marker Off softkey in the marker menu of the base unit.
Percent Marker
"Opens an edit dialog box to enter a probability value and to position marker 1. Thus, the power
which is exceeded with a given probability can be determined very easily. If marker 1 is
deactivated, it will be switched on automatically.
This softkey is only available for the CCDF measurement.
Remote: CALC:MARK:Y:PERC 0…100%