EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 766

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSL Common Commands
1300.2519.12 6.7 E-11
*ESR?
STANDARD EVENT STATUS QUERY returns the contents of the event status register in
decimal form (0 to 255) and subsequently sets the register to zero.
Parameter
0 to 255
*IDN?
IDENTIFICATION QUERY queries the instrument identification.
Return values (examples)
LEGacy format, example for R&S FSL6:
Rohde&Schwarz,FSL-6,100005/016,1.80
NEW format, example for R&S FSL6:
Rohde&Schwarz,FSL-6,1300.2502K16/100005,1.80
FSL-6 device name
100005/016 serial number/model
1300.2502K16/100005 part number/serial number
1.80 firmware version
*IST?
INDIVIDUAL STATUS QUERY returns the contents of the IST flag in decimal form. The IST flag
is the status bit which is sent during a parallel poll (see chapter 5 "Remote Control - Basics").
Parameter
0 | 1
*OPC
OPERATION COMPLETE sets bit 0 in the event status register after all preceding commands
have been executed. This bit can be used to initiate a service request (see chapter 5 "Remote
Control - Basics").
*OPT?
OPTION IDENTIFICATION QUERY queries the options included in the instrument and returns a
list of the options installed. The options are separated from each other by means of commas.
Parameter
B<number> hardware options
K<number> software options
For a list of all available options and their description refer to the CD-ROM.
Example
B4,B5,B6,B7,B8,B10,B22,B30,B31,K7,K9

Table of Contents

Related product manuals