CALCulate Subsystem R&S FSL
1300.2519.12 6.36 E-11
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:LIM:ACP:ACH:RES?
Queries the limit check result in the adjacent channels.
Characteristics
*RST value: OFF
SCPI: device–specific
Mode
A, CDMA, EVDO
CALCulate<1|2>:LIMit<1...8>:ACPower:ACHannel:RESult?
This command queries the result of the limit check for the upper /lower adjacent channel when
adjacent channel power measurement is performed.
If the power measurement of the adjacent channel is switched off, the command produces a
query error.
The numeric suffixes <1|2> and <1...8> are irrelevant for this command.
This command is a query and therefore has no *RST value.
Parameter
The result is returned in the form <result>, <result> where <result> = PASSED | FAILED, and
where the first returned value denotes the lower, the second denotes the upper adjacent
channel.
Example
CALC:LIM:ACP:ACH 30DB, 30DB
Sets the relative limit value for the power in the lower and upper adjacent channel to 30 dB
below the channel power.
CALC:LIM:ACP:ACH:ABS –35DBM, –35DBM
Sets the absolute limit value for the power in the lower and upper adjacent channel to –35 dB.
CALC:LIM:ACP ON
Switches on globally the limit check for the channel/adjacent channel measurement.
CALC:LIM:ACP:ACH:STAT ON
Switches on the limit check for the adjacent channels.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:LIM:ACP:ACH:RES?
Queries the limit check result in the adjacent channels.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A, CDMA, EVDO, WLAN