CALCulate Subsystem R&S FSL
1300.2519.12 6.78 E-11
CALCulate<1|2>:MARKer<1...4>:FUNCtion:NDBDown:TIME?
This command queries the values of the two temporary markers in zero span. The time values
are separated by comma and output in ascending order.
A complete sweep with synchronization to sweep end must be performed between switching on
the function and querying the measured value to obtain a correct query result. This is only
possible in single sweep mode.
The numeric suffixes <1|2> and <1...4> are irrelevant for this command.
This command is only a query and therefore has no *RST value.
Example
INIT:CONT OFF
Switches to single sweep mode
CALC:MARK:FUNC:NDBD ON
Switches on the n dB down function.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:NDBD:TIME?
Outputs the time values of the temporary markers.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A–T
CALCulate<1|2>:MARKer<1...4>:FUNCtion:NOISe[:STATe]
This command switches the noise measurement on or off for all markers. The noise power
density is measured at the position of the markers. The result can be queried with
CALCulate<1|2>:MARKer<1...4>:FUNCtion:NOISe:RESult?.
The numeric suffixes <1|2> are irrelevant for this command.
Parameter
ON | OFF
Example
CALC:MARK:FUNC:NOIS ON
Switches on the noise measurement.
Characteristics
*RST value: OFF
SCPI: device–specific
Mode
A