R&S FSL CALCulate Subsystem
1300.2519.12 6.105 E-11
Characteristics
*RST– value: –
SCPI: device–specific
Mode
A–T
CALCulate<1|2>:MARKer:FUNCtion:SUMMary:RMS:RESult?
This command queries the result of the measurement of the RMS power value.
A complete sweep with synchronization to sweep end must be performed between switching on
the function and querying the measured value to obtain a correct query result. This is only
possible in single sweep mode.
The numeric suffixes <1|2> are not relevant.
This command is only a query and therefore has no *RST value.
Example
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:SUMM:RMS ON
Switches on the function.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:SUMM:RMS:RES?
Outputs the result.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A–T
CALCulate<1|2>:MARKer:FUNCtion:SUMMary:SDEViation[:STATe]
This command switches on or off the measurement of the standard deviation.
On switching on the measurement, the mean power measurement is switched on as well.
The numeric suffixes <1|2> are not relevant.
Parameter
ON | OFF
Example
CALC:MARK:FUNC:SUMM:SDEV ON
Switches on the measurement of the standard deviation.
Characteristics
*RST value: OFF
SCPI: device–specific
Mode
A