R&S FSL SENSe Subsystem
1300.2519.12 6.177 E-11
Example
BAND:TYPE NORM
Characteristics
*RST value: NORMal
SCPI: device–specific
Mode
all, except ADEMOD
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo
This command defines the instruments video bandwidth. The available video bandwidths are
specified in the data sheet. The command is not available if FFT filtering is switched on and the
set bandwidth is 30 kHz or if the quasi–peak detector is switched on.
The numeric suffixes <1|2> are not relevant.
Parameter
refer to data sheet
Example
BAND:VID 10kHz
Characteristics
*RST value: – (AUTO is set to ON)
SCPI: conform
Mode
A, CATV, EVDO
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo:AUTO
This command either automatically couples the instruments video bandwidth to the resolution
bandwidth or cancels the coupling.
The ratio video bandwidth/resolution bandwidth can be modified with the
[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]:RATio command.
The numeric suffixes <1|2> are not relevant.
Parameter
ON | OFF
Example
BAND:VID:AUTO OFF
Characteristics
*RST value: ON
SCPI: conform
Mode