6-1
IM 701210-05E
Triggering
6
Chapter 6 Triggering
6.1 Setting the Trigger Mode
<For a description of this function, refer to page 2-27.>
Procedure
MEASURE
CURSOR RESET SELECT
CH
ALL CH
MODE
POSITION
SIMPLE/ENHANCED
ACQ
START/STOP
SETUP DISPLAY
ZOOM
DUAL
CAPTURE
HISTORY
MATH
1
CH
2
CH
3
CH
4
CH
5
CH
6
CH
7
CH
8
CH
9
CH
10
CH
11
CH
12
CH
13
CH
14
CH
15
CH
16
DELAY
MANUAL TRIG
V/DIV TIME/DIV
TRIGGER
TRIG D
VERTICAL
HORIZONTAL
CAL
X-Y
SEARCH
7
DSP 1
8 9
DSP 2
654
1
0
23
ENTER
m
DSP 6DSP 5
LOGIC A
LOGIC B
EVENT
DSP 3 DSP 4
EXP
GO/NO-GO
ACTION
FILE
SHIFT
1. Press MODE.
2. Press the soft key corresponding to the desired mode to set the trigger mode.
3. If you select Single(N), turn the jog shuttle to set Single(N) Count (the number
of times to acquire the waveform).
Explanation
Auto Mode
If the trigger condition is met within the 50-ms timeout period, the waveform is updated
on each trigger occurrence. If the trigger condition is not met after the timeout period
elapses, the waveform is automatically updated. If the time axis setting is in the range in
which the display mode is set to roll mode, the display is set to roll mode.
However, triggering operates in normal mode when using simple trigger with the trigger
source set to Time, even if auto mode is specified.
Auto Level Mode
Within the timeout period (approximately 1 s), the waveforms are displayed in the same
fashion as in the auto mode. If a trigger does not occur within the timeout period
(approximately 1 s), the center value of the amplitude of the trigger source is detected,
the trigger level is automatically changed to the center value, and the trigger is generated
to update the displayed waveform. The auto-level mode is valid only if the trigger is a
simple trigger and the trigger source is between CH1 and CH16. For all other cases, the
operation is the same as the auto mode.
If the time axis setting is in the range in which the display mode is set to roll mode, the
display is set to roll mode.