7-1
IM 701210-05E
Acquisition and Display
7
Chapter 7 Acquisition and Display
7.1 Starting/Stopping Waveform Acquisition
Procedure
MEASURE
CURSOR RESET SELECT
ALL CH
ACQ
START/STOP
SETUP DISPLAY
ZOOM
DUAL
CAPTURE
HISTORY
MATH
CAL
X-Y
SEARCH
GO/NO-GO
FILE
SHIFT
HELP
MISC
NUM KEY
Setting the Waveform Acquisition Mode at Power ON
1. Press MISC.
2. Press the Next 1/2 soft key
3. Press the Others soft key.
4. Press the Start Mode(Power On) soft key to select ON or OFF.
START/STOP Key Response Time Setting
5. Press the START/STOP Response Time soft key, then select Quick or >1sec.
Starting/Stopping Waveform Acquisition
6. Press START/STOP. Waveform acquisition is started/stopped.
Waveform acquisition is in progress when the indicator above and to the right of
the key is illuminated.
Explanation
Waveform Acquisition and Indicator Display
• If the indicator above START/STOP is illuminated, the waveform acquisition is in
progress. “Running” is displayed at the lower left corner of the screen.
• If the indicator above START/STOP is not illuminated, the waveform acquisition is
stopped. “Stopped” is displayed at the lower left corner of the screen.
Operation When the Acquisition Mode Is Set to Averaging Mode
• Averaging is stopped when waveform acquisition is stopped.
• Averaging starts again when waveform acquisition is restarted.
START/STOP Operation during Accumulation
Accumulation is suspended when acquisition is stopped.
It is resumed when acquisition is restarted.