6-7
IM 701210-05E
Triggering
6
Selectable Range of Hold Off Time
0 ns to 10 s (the initial value is 0 ns). The resolution is 100 ns.
Notes for Setting the Hold Off Time
If you are setting the hold off time to 50 ms or greater, set the trigger mode to normal.
When used with A->B(N) or A Delay B trigger, the hold off time applies only to
condition B.
The hold off time is set to 0 ns for Period trigger, and this function is invalid.