EasyManua.ls Logo

YOKOGAWA DL750 - Page 243

YOKOGAWA DL750
747 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-25
IM 701210-05E
Triggering
6
Explanation
This function activates a trigger on the OR logic of the edge trigger of each channel while
condition A is true.
Setting Conditions A/Edge Or
Pattern of Each Channel of Condition A: CH1 to CH16, Logic A, and Logic B
Select from the following:
CH1 to CH16 (Other Than Logic Inputs)
H: Above the preset trigger level
L: Below the preset trigger level
X: Don’t Care
Logic Input
Enable: Make the combination of the pattern
1
of each bit the trigger condition
Disable: Don’t Care
1. Select the pattern of each bit from the following:
H: Above a certain level
2
L: Below a certain level
2
X: Don’t Care
2. Varies depending on the logic probe being used as follows:
700986: Approx. 1.4 V
700987: 6 V± 50% (for DC input)
700987: 50 V± 50% (for AC input)
Condition
Select from the following:
True: A trigger is activated while all channels match the specified pattern.
False: A trigger is activated while any of the channels do not match the specified
pattern.
Edge Trigger of Each Channel
This setting is possible only if the channel’s condition A pattern is set to X (Don’t
Care). If you set all condition A patterns to X (Don’t Care), the trigger will operate as
an OR trigger.
CH1 to CH16 (Other Than Logic Inputs)
: Rising edge
: Falling edge
–: Don’t Care
Logic Input
Enable: Make the combination of the pattern
1
of each bit the trigger condition
Disable: Don’t Care
1. Select the pattern of each bit from the following:
: When the signal goes above a certain level
2
: When the signal goes below a certain level
2
–: Don’t Care
2. Varies depending on the logic probe being used as follows:
700986: Approx. 1.4 V
700987: 6 V± 50% (for DC input)
700987: 50 V± 50% (for AC input)
6.12 Setting the Edge on A Trigger (ENHANCED)

Table of Contents