16-36
IM 701210-06E
Setting the Display Format (Display)
Display Format
Select the number of divided windows when displaying waveforms from the
following:
Single, Dual, Triad, Quad, Octal, and Hexadecimal
For the meanings of the selections, see the explanation in section 8.1.
Turning ON/OFF the Display of the Waveform Labels and Scale Values
(Label)
• Display Waveform Label (TraceLabel)
To display the labels of the displayed waveforms, select the TraceLabel check
box. Clear the check box to not display the labels.
• Turning ON/OFF the Display of the Upper and Lower Limits of the Displayed
Waveforms (ScaleValue)
To display the upper and lower limits of the displayed waveforms, select the
ScaleValue check box. Clear the check box to not display the upper and lower
limits.
Selecting the Waveforms to Be Saved (File)
Selecting the Target Waveforms (Trace)
From the Trace box, select the waveforms to be saved from the following:
All (all waveforms), 1 to 16 (channels 1 to 16), DSP1 to DSP6 (optional), Math 1 to
Math 8 (computed waveforms), PodA to PodB (logic waveforms), and Event (event
waveform).
Selecting the Save Area (Range)
Select the area in which the target waveform to be saved is displayed from the
following:
Main, Z1, and Z2
For the meanings of the selections, see the explanation in section 13.7.
Setting the History Data Display
Selecting by History Data Number (History No.)
You can specify the history number of the data to be displayed.
Selecting the Newest or Oldest Data
You can specify the newest or the oldest data in the history data to be displayed.
Selecting One or All
• One
Click One to display a single history data point that is selected by the data
number or selected by newest or oldest.
• All
Click All to display all the history data.
Updating the Setup Condition (Panel Update)
Click Panel Update to update the control panel settings on the Data Capture window
(PC) to match the newest setup condition on the DL750/DL750P.
16.7 Using the Web Server Function