App-6
IM 701210-06E
Possible Time Axis Range, Maximum Record Length, and Maximum Sample Rate for
Realtime Recording
Varies depending on the number of channels that are to be realtime recorded as follows:
Channels Used Time Axis Range Maximum Maximum
Record Length Sample Rate
18CH or more 20 s/div to 3 day/div 25 MW 5 kS/s
12 to 17CH 20 s/div to 3 day/div 50 MW 5 kS/s
6 to 11CH 10 s/div to 3 day/div 50 MW 10 kS/s
4 to 5CH 6 s/div to 3 day/div 100 MW 20 kS/s
3CH 3 s/div to 3 day/div 250 MW 50 kS/s
2CH 1 s/div to 3 day/div 500 MW 100 kS/s
1CH 1 s/div to 3 day/div 1 GW 100 kS/s
Appendix 2 Relationship between the Record Length and Acquisition Mode