2-58
IM 701210-05E
GO/NO-GO Determination <Sections 11.8 to 11.10>
The GO/NO-GO function is useful when you want to inspect signals and track down
abnormal symptoms on a production line making electronic equipment. The function
determines whether the waveform is within the preset range and performs a specific
action for a GO or NO-GO result. There are two types of determinations.
• Zone: Method in which a waveform zone is set on the screen
• Parameter: Method in which a waveform parameter range is specified
The following five types of actions can be performed for GO or NO-GO result.
• Print the Screen Image Data (PRINT)
Prints the screen image data to a specified printer.
• Save the Screen Image Data (Image)
Saves the screen image data to the save destination specified in the FILE menu.
• Save the Waveform Data (Save to File)
Saves the waveform data in binary, ASCII, or floating format to the save destination
specified in the FILE menu.
• Beep Sound (Buzzer)
Sounds a buzzer.
• Send Mail
Sends an e-mail message to a specified address. (when the Ethernet interface option
is installed)
In addition, the determination result can be output externally using signals.
2.7 Waveform Analysis/Search

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL750 and is the answer not in the manual?

YOKOGAWA DL750 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL750
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals