3. Repair Instructions 1/2010 3 - 67
BA-TE-DE08C M.KAY
Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
Field Service Utilities V2.00 - SetMST
CRC = XXXX - OK
Write Calibration Data (USB Stick - > Dialog+ Hard Disk)
No source is found...
AQ
: back to previous menu
The utility searches for sources. An input screen can
open with the following information (see left figure),
i.e. no sources were found on the USB stick because the
calibration data subdirectories are empty. Calibration
data must be read to the USB stick before you can
write calibration data to the Dialog+ hard disk drive
(CFC).
3.
Press
AQ
to go back to the previous menu.
4.
Use the
Read Calibration Data
utility to read the
calibration data from the Dialog+ hard disk drive
(CFC) to the USB stick.
Field Service Utilities V2.00 - SetMST
CRC = XXXX - OK
Write Calibration Data (USB Stick - > Dialog+ Hard Disk)
Writing Calibration Data to Dialog+ from: Ser-No./SYSver=90832/VX_XX
Are you sure?
AQ
: no
EQ
: yes
An input screen opens to
Write Calibration Data
to the
hard disk drive (CFC) with the following hardware keys:
AQ
: No, calibration data are not written to
Dialog+ hard disk drive (CFC), exit to main
menu
EQ
:
Yes, calibration data are written to Dialog+
hard disk drive (CFC)
Ser-No./SYSver=90832/VX_XX
The utility compares the serial number (Ser-No.) and the system version number
(SYSver) stored in the calibration data file with the serial number and the
system version number from the Dialog+. The calibration data file can be
written only if they match.
Example
Ser-No.= 90832
SYSver = V8_20