4. TSM Service Program 1/2010 4 - 33
BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd B. Braun Avitum AG
Dialog
+
SW 9.xx
4.7.1.7 Rinsing Parameters
The Rinsing Parameters can be set in the Rinsing
Parameters menu.
1. Activate the System Configuration menu with
the System Configuration key in menu
Treatment Support. The following menu is
opened and displayed (scroll with scroll bar).
2. Activate the Rinsing Parameters menu with
the Rinsing Parameters icon. The following
menu is displayed.
The following parameters can be set:
Priming without Recirculation
Filling BP Rate
Filling BP Volume
Automatic Rinse Program after Self Tests
User Message for BS Filled?
Rinsing Blood Flow
Rinsing Dialysate Flow
Rinsing Time by UFP
Rinsing Rate by UFP
Rinsing Volume by UFP
Blood Flow for Connecting Patient