EasyManua.ls Logo

B. Braun Dialog+ - SAD Test after Calibration

B. Braun Dialog+
455 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4. TSM Service Program 1/2010 4 - 150
BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.3.12.1 SAD Test after Calibration
SAD Test
AIR must be displayed if the SAD TEST is activated, even if a tubing filled with
fluid is inserted in the SAD.
The displayed result with SAD TEST OFF depends on the tubing inserted in the
SAD:
AIR: must be detected if a tubing filled with air is inserted.
NO AIR: must be detected if a tubing filled with fluid is inserted.
1. Activate the menu Test 1.7 Air Sensor, Red
Sensor in the LL Manual Test menu.
2. Insert a tubing filled with water in the SAD
and close cover.
The following settings must be displayed:
SAD LEVEL:
15
(Desired Setting)
SAD TEST:
NO AIR
(Controller/Supervisor)
RDV:
NO BLOOD
(Controller/Supervisor)
3. Press SAD TEST key (Desired
Setting).
The following settings must be displayed:
SAD LEVEL:
KS+5
(Desired Setting, see previous
page)
SAD TEST:
TEST
(Desired Setting)
SAD TEST:
AIR
(Controller/Supervisor)
4. Press the SAD TEST key (Desired
Setting) to terminate the test.
The following setting must be displayed:
SAD TEST:
OFF
(Desired Setting)
5. Press the SAD TRIGGER OFF key
(Desired Setting) to activate the SAD
TimeContr.
The following settings must be displayed:
SAD TimeContr.
:
OFF
(Supervisor)
after approx. 1 second:
SAD TimeContr.:
ON
(Supervisor)
6. Remove the tubing from the SAD.
7. Interrupt the light path in the SAD with a
piece of paper.
The following settings must be displayed:
RDV:
NO BLOOD
(Controller/Supervisor)
changes to
RDV:
BLOOD
(Controller/Supervisor)

Table of Contents

Related product manuals