3. Repair Instructions 1/2010 3 - 8
BA-TE-DE08C M.KAY
Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
3.2.3 Example LLC/LLS Test Codes
for Blood Leak Detector
LLC Test Code:
11 (TT):
Blood leak detector test
10 (SS):
Send signal to blood leak task
2 (x):
Third repetition (x = 0, 1, 2)
LLS Test Code:
11 (TT):
Blood leak detector test
004 (SS):
Measure level for red and green
LLC Test
Code
LLC Test Steps LLS Test Steps LLS Test
Code
1.
1102x
Prerequisites for Test:
Upline tank must be filled
Timeout 2 min
2.
Test code =
11
to LLS Test code =
11
back to LLC
3.
1110x
EV_TESTE_BLUTLECK_EIN to BlutleckTaskID Test initialisation (VGD ON)
Single measurement of green,
Single measurement of red,
then comparison of: red
green + 0.2 V
no timeout
11000 –
11003
11004
4.
Acknowledge with switch-off of VGD 11005 –
5.
1120x Wait for D24OFF (timeout 30 s) 11011
6.
VGD ON 11012
7.
1130x Wait for D24ON (timeout 30 s)
8.
1199x Completed
3.2.4 Overview Test Codes
00
No tests
11
Blood leak detector
12
DFS pressure test
15
Integrity test HDF online (only for Dialog+ HDF online)
16
Alarm system
17
Monitoring of analog voltage
±
12 V
20
UF pump
21
End conductivity
22
Temperature TSD
25
SMPS-MC EEPROM
26
SMPS-MC service mode
27
SMPS-MC battery
28
SMPS-MC buzzer
30
SAD: Level test
31
SAD: Counter test
32
Pressure sensors blood side: Equality test for blood side pressure sensors
33
Pressure sensors blood side: Blood side pressure test
37
Disinfection Valve VD
40
Pressure test substitution line S-Online HDF online
11102
11004

Table of Contents