4. TSM Service Program 1/2010 4 - 39
BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd B. Braun Avitum AG
Dialog
+
SW 9.xx
4.7.1.10 Disinfection Parameters
Disinfection after servicing in TSM
Before disinfection: rinse the machine to remove residual dialysate or
concentrate in the system.
The disinfection parameters can be set in the
Disinfection Parameters menu.
1. Activate the System Configuration menu with
the System Configuration key in menu
Treatment Support. The following menu is
opened and displayed (scroll with scroll bar).
2. Activate the Disinfection menu with the
Disinfection Parameters icon. The following
menu is displayed.
3. Scroll the menu with the scroll bar.
The following parameters can be set:
Disinfection after each Therapy
Automatic Preparation Start after Disinfection
Thermal Disinfection
Chemical Disinfection
Chemical Disinfection, Short
Central Thermal Disinfection
Central Manual Chemical Disinfection
Central Automatic Chemical Disinfection
Rinsing
Auto Switch-On with Rinsing (No Weekly
Schedule)
- Switch-On Time
Auto Switch-On by Weekly Schedule
- Weekly Schedule
- Start on Sunday/Start on Monday
Auto Switch-On without Daily Confirmation
Maximum Out of Action Time (Day/Hour/
Minutes)
Disinfection Configuration Data
Chemical (long), Chemical (short)
Disinfection Configuration Data
Central Thermal, Central Chemical, Rinsing