EasyManua.ls Logo

B. Braun Dialog+ - Miscellaneous Parameters

B. Braun Dialog+
455 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4. TSM Service Program 1/2010 4 - 54
BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd B. Braun Avitum AG
Dialog
+
SW 9.xx
4.7.1.19 Miscellaneous Parameters
Further parameters can be set in the
Miscellaneous Parameters menu.
1. Activate the System Configuration menu with
the System Configuration key in menu
Treatment Support. The following menu is
opened and displayed (scroll with scroll bar).
2. Activate the Miscellaneous Parameters menu
with the Miscellaneous Parameters icon. The
following menu is displayed.
The following parameters can be set:
Click Sound
Warning and Writing (Saving) Patient
Parameters
Skip Self Tests
Automatic Preparation Start
Data Validation with
UF Profile Editing in Treatment
Manual Bypass Confirmation Window
End of Therapy Sound Duration
Chopped Alarm Sound
Suppression of Warning Sounds in Preparation
Automatic Reinfusion Start at Entering End of
Treatment
Click Sound
If you activate the YES key a click sound is activated if a key or the touch
screen is pressed.
Warning and Writing (Saving) Patient
Parameters
If you click the YES key you can generate a warning message to save the
patient data in the following cases:
The patient data was loaded from the data media
The therapy was terminated without saving the patient data
Skip Self Tests It is only permitted
to activate the
Skip Self Test
function for servicing.
It is prohibited
to run a therapy with a patient if the
Skip Self Test
function is activated.
The self test must be
passed successfully prior to a therapy.

Table of Contents

Related product manuals