4. TSM Service Program 1/2010 4 - 88
BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.2.14 Test 1.14 Bypass and Disinfection Valves
1. Activate the Bypass and Disinfection Valves
menu with the Bypass and Disinfection Valves
key in the LL Manual Test menu. The following
menu is displayed.
2. Check the functions with the appropriate keys.
You can activate a single valve or combined setting via the controller. The
actual condition is displayed by the supervisor. The following valves can be
activated:
Valves VBP Bypass valve
VDE Dialyser inlet valve
VDA Dialyser outlet valve
VD Disinfection valve
VZ Circulation valve
VLA Air separator valve