4. TSM Service Program 1/2010 4 - 36
BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd B. Braun Avitum AG
Dialog
+
SW 9.xx
Gross UF Rate vs. Blood Flow -
Warning Limit
You can set and activate a warning limit, i.e. a warning (yellow) is activated if
the limit is reached.
Setting Range
The alarm limit can not be set lower than the warning limit (even if the
warning limit is deactivated).
There are no limit alarms for HDF machines in predilution.
Alarm Limit/Warning Limit Alarm Limit Warning Limit
deactivated deactivated No alarm/no warning in therapy
deactivated activated Warning in therapy if limit is
exceeded
activated deactivated Alarm in therapy if limit is exceeded
activated activated Alarm and warning in therapy if
limit is exceeded
UF Profile Editor
The UF profile can be edited in the UF Profile Editor. Editing can be performed
in the therapy mode and in the TSM service program.
Press the UF Profile Editor icon to open the UF
Profile Editor menu.
CONST (Constant)
If this key is activated all intervals of the profile
are set to 100%.
LIN (Linear)
If this key is activated only the first interval can
be set. A linear adaptation is calculated for the
remaining process.
EXP (Exponential)
If this key is activated only the first interval can
be set. An exponential adaptation is calculated for
the remaining process.