4. TSM Service Program 1/2010 4 - 47
BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd B. Braun Avitum AG
Dialog
+
SW 9.xx
Height Measurement Unit of
Watson Formula
The unit [cm] or [Feet/Inch] can be selected.
Default Setting for Treatment Mode:
Warning of Kt/V Target Projected Deviation
If the calculated Kt/V from the set parameters is smaller than the set target
value a message is activated.
Kt/V Table - Manual INPUT:
Setting Clean Blood Volume [%] Enabled
If YES is Activated
The ratio can be edited subsequently in therapy for adaptation to the
calculated Kt/V value.
Warning for Saving Kt/V Table
A warning message can be enabled for saving the Kt/V table.
Dialyser Filter Urea Clearance Data
The Kt/V is approximately calculated via the
Dialyser Filter Urea Clearance Data for the
selected dialyser (see table, use scroll bar to see
table).
Details can be found in Chapter 3 repair
instructions.