EasyManua.ls Logo

B. Braun Dialog+ - Page 349

B. Braun Dialog+
455 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5. Technical Safety Inspection and Preventive Maintenance 1/2010 5 - 4
BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd B. Braun Avitum AG
Dialog
+
SW 9.xx
S
M
Check List (S = Technical Safety Inspection Points; M = Preventive Maintenance Points)
Note: Text in { } brackets is information for the execution of the check list!
NO YES OK
M
2.22 Conductivity Sensors
2.22.1 ENDLF sensor: Visual inspection
2.22.2 BICLF sensor: Visual inspection
M
2.23 Piston Pumps
2.23.1 BIC piston pump: Visual inspection
2.23.2 KP piston pump: Visual inspection
M
2.23.3 UF piston pump: Visual inspection
M
2.24 Blood leak detector (BLD) test passed
2.24.1
Check blood leak detector and calibrate if necessary
n
M
2.25
Degassing Pressure Sensor PE (permissible tolerance ±13 [mbar])
- Comparison measurement (at approx. -530): Measurement instrument: ......................... Display on machine: ......................... [mbar]
- Comparison measurement (at approx. -130): Measurement instrument: ......................... Display on machine: ......................... [mbar]
M
2.26
Dialysate Outlet Pressure Sensor PDA (permissible tolerance ±13 [mbar])
- Comparison measurement (at approx. -530): Measurement instrument: ......................... Display on machine: ......................... [mbar]
- Comparison measurement (at approx. 0): Measurement instrument: ......................... Display on machine: ......................... [mbar]
- Comparison measurement (at approx. +530): Measurement instrument: ......................... Display on machine: ......................... [mbar]
M
2.27
Heparin Pump
2.27.1
Position (80 mm ± 15 mm) {Drive fast forward to end position with 30 ml syringe. Check display position in mm.} ......................... [mm]
M
2.28 Set summer/winter time {TSM, Treatment Support, System Configuration, if applicable in this country}
3. Function Inspection (Document Measurement Values)
S
3.1
Blood Pump: - Check function, moveability, noise rating
3.1.1 - Check roller/one-way bearing; clean and lubricate shaft slightly if necessary
3.1.2 - Alarm cover switch
S
3.2
Single-Needle Pump:
Option present
3.2.1 - Check function, moveability, noise rating
3.2.2 - Check roller/one-way bearing; clean and lubricate shaft slightly if necessary
3.2.3 - Alarm cover switch
S
3.3
Venous Tubing Clamp SAKV: - Function/gap checked, set gap 1.4 (+0.1 mm) if necessary
S
3.4
Arterial Tubing Clamp SAKA:
Option present
3.4.1 - Function/gap checked, set gap 1.4 (+0.1 mm) if necessary
S
3.5
Arterial Pressure PA: (permissible. tolerance ±10 [mmHg])
Comparison measurement (at approx. -400): Measurement instrument: ......................... Display on machine: ...................... [mmHg]
Comparison measurement (at approx. 0): Measurement instrument: ......................... Display on machine: ...................... [mmHg]
S
3.6
Inlet Pressure PBE: (permissible. tolerance ±10 [mmHg])
Comparison measurement (at approx. 0): Measurement instrument: ......................... Display on machine: ...................... [mmHg]
Comparison measurement (at approx. +400): Measurement instrument: ......................... Display on machine: ...................... [mmHg]
S
3.7
Venous Pressure PV: (permissible. tolerance ±10 [mmHg])
- Comparison measurement (at approx. 0): Measurement instrument: ......................... Display on machine: ...................... [mmHg]
Comparison measurement (at approx. +400): Measurement instrument: ......................... Display on machine: ...................... [mmHg]
S
3.8
Single-Needle Pressure PBS: (permissible. tolerance ±10 [mmHg])
Option present
- Comparison measurement (at approx. 0): Measurement instrument: ......................... Display on machine: ...................... [mmHg]
- Comparison measurement (at approx. +400): Measurement instrument: ......................... Display on machine: ...................... [mmHg]

Table of Contents